This chapter describes Dionex AutoTrace 280 service and repair procedures that
users can perform. All procedures not included here, including electronics-related
repair procedures, must be performed by Thermo Fisher Scientific personnel. For
assistance, contact Technical Support for Dionex products:
In the U.S. and Canada, call 1-800-532-4752 and select option 2.
Outside the U.S. and Canada, call the nearest Thermo Fisher Scientific office.
Please have the following information at hand when talking with the Technical
Support Representative:
Dionex AutoTrace 280 serial number (printed on the model data label)
AutoTrace software version (select Help > About to display the version)
Before replacing any part, see the troubleshooting information in
correctly identify the cause of the problem.
Substituting non-Dionex/Thermo Fisher Scientific parts may impair
the performance of the Dionex AutoTrace 280, thereby voiding the
product warranty. For details, see the warranty statement in the
Dionex Terms and Conditions.
5.1

Replacing Tubing

Replace tubing assemblies whenever tubing becomes kinked or develops a
blockage that prevents flow. If tightening a fitting that is leaking does not resolve
the problem, the fitting or the entire tubing assembly should be replaced.
All replacement tubing must have the same length and ID as the
original tubing. This maintains the accuracy of the Dionex AutoTrace
280 liquid and air push operations.
Parts and Tools Needed
#2 Phillips screwdriver
Replacement tubing (for the standard Dionex AutoTrace 280, see
for the Dionex AutoTrace 280 PFAS, see
Doc. 065330-05 03/2020