Endress+Hauser CCS50D Operating Instructions Manual page 30

Digital sensor with memosens technology for determining chlorine dioxide
Table of Contents

Advertisement

Diagnostics and troubleshooting
Error
Display value too low
Display fluctuates
considerably
30
Possible cause
Membrane cap not screwed on fully
Membrane soiled
Air bubble in front of membrane
Air bubble between cathode and
membrane
Input flow of medium too low
Foreign oxidants interfering with
DPD reference measurement
Use of organic disinfectants
Hole in membrane
Remedy
Fill membrane cap with fresh electrolyte
→  34
Screw membrane cap on fully
Clean membrane →  32
Release air bubble
Remove membrane cap, top up electrolyte
Remove air bubble by tapping on the
outside of the membrane cap
Screw on membrane cap
Establish correct flow
Examine medium, check chemicals
Use suitable agent (e.g. as per DIN 19643)
(water may need to be replaced first)
Use suitable reference system
Replace membrane cap
CCS50D