3.9 Install the titrant and the reagent

Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to the
current safety data sheets (MSDS/SDS) for safety protocols.
Loosen the tube connector on the bottle cap.
Put a filled desiccant cartridge into the adapter on the titrant and solvent bottle caps. Put
an empty desiccant cartridge into the adapter on the waste bottle cap.
Note: The molecular sieve in the filled cartridges must be regenerated regularly depending on the
humidity of the laboratory. Refer to
Push the inlet tube through the connector. Make sure that the end of the tube is at the
bottom of the bottle. Tighten the connector on the bottle cap.
Pump 1 is used to fill the measurement cell with solvent. Pump 2 is used to empty the
measurement cell. Refer to the illustrated steps that follow.
C A U T I O N
Regenerate the molecular sieve
Installation
on page 30.
13