7.7 Storage and transportation

7.7.1 Prepare the instrument for short-term storage

7.7.2 Prepare the instrument for long-term storage

7.7.3 Prepare the instrument for shipment

Option
Maintenance
summary
Maintenance
schedule
Reagent
replacement
1. Push the power button for 2 seconds to set the power to off. The contents of the
syringe drain back into the titrant bottle.
2. Drain the used beakers, then clean the beakers with deionized water.
3. Rinse the sensor with deionized water. Put the sensor in a storage tube that contains
a storage solution. Refer to the sensor documentation for storage instructions.
1. Push the power button for 2 seconds to set the power to off. The contents of the
syringe drain back into the titrant bottle.
2. Drain the used beakers, then clean the beakers with deionized water.
3. Remove the tubes from the titrant bottles and the reagent bottles. Seal the bottles
with the original bottle caps.
4. Flush the syringe and pump tubes with deionized water or an applicable cleaning
solution.
5. Rinse the sensor with deionized water. Put the sensor in a storage tube that holds a
storage solution. Refer to the sensor documentation for storage instructions.
1. Refer to
Prepare the instrument for long-term storage
2. Put the instrument and the necessary accessories in a package with all the packaging
materials.
3. Ship the instrument at a temperature –5 to 40 °C (23 to 104 °F) and relative humidity
up to 80%.
4. Make sure that the storage location for the instrument does not have dust,
condensation or chemical evaporation.
Description
See the number of days remaining for maintenance tasks. After doing a
task, push Reset to set the number of days remaining to the default
value.
See the list of maintenance tasks. Push Edit to change the default
value.
Use this option to replace the reagents. Obey the instructions on the
display.
Maintenance
on page 33.