Section 3

3.1 Installation guidelines

3.2 Connect to AC power

3.3 Install the syringe

Installation
Multiple hazards. Only qualified personnel must conduct the tasks described in this
section of the document.
This instrument is for indoor use only.
The power supply connector on the rear panel must be easily accessible so the
power can be disconnected quickly in case of emergency.
Keep the instrument away from temperature extremes, including heaters, direct
sunlight and other heat sources.
Put the instrument on a stable and level surface in a well ventilated place.
Make sure that there is at least 15 cm (6 in.) of space on all sides of the instrument to
prevent electrical parts from overheating.
Do not operate or keep the instrument in dusty, damp or wet locations.
Always keep the surface of the instrument and all accessories dry and clean.
Electrocution hazard. If this equipment is used outdoors or in potentially wet locations, a
Ground Fault Circuit Interrupt (GFCI/GFI) device must be used for connecting the
equipment to its main power source.
Electrical shock and fire hazards. Make sure that the supplied cord and non‐locking plug
meet the applicable country code requirements.
Fire hazard. Use only the power supply that is specified for this instrument.
1. Connect the power cord to the power supply.
2. Connect the power supply to the instrument. Refer to
3. Connect the power cord to an electrical outlet.
Before syringe installation, set the instrument power to on. Push the power button on the
front of the instrument. Make sure that the startup sequence shows on the display. The
syringe holder lowers to its operating position.
Note: Ignore any warning messages related to missing applications that show on the display.
The sensor holder has two positions: one over the magnetic stirrer and the second at
180° to the right. Move the sensor holder away from the instrument to the second
position.
Refer to the illustrated steps that follow.
C A U T I O N
D A N G E R
C A U T I O N
W A R N I N G
Figure 2
on page 7.