Section 5

5.1 Configure the instrument

5.1.1 Change the application settings

Startup
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to the
current safety data sheets (MSDS/SDS) for safety protocols.
Personal injury hazard. Never use the instrument without the syringe cover installed.
1. From the main menu, select
2. Select an option, then push Select.
Option
Description
Applications
Exports, changes (refer to
removes and makes copies of application data. Make sure that the
duplication function does not make more than five applications.
Operators
Adds, changes and removes operators.
Date + Time
Sets the instrument date and time.
Brightness
Sets the brightness of the display.
Sounds
Sets the sound options.
Language
Sets the language.
Network
Give a name to the instrument. This name is used to connect the
instrument to a PC. Restart the instrument if the name is changed.
Legacy settings
Specify the sensor data when the legacy adapter is used.
Info
Shows information about the instrument and the attached hardware.
Restore Defaults Sets the instrument to the default configuration.
Options
Sets the application parameters view to expert mode (refer to
application settings
the syringe to empty into the titrant bottle. Changes the temperature
display unit. Prints the measurement curve if a printer is connected.
Note: When the instrument is set to on, the application parameters view is
always set to basic mode. Changes made to application data with the
instrument in expert mode are kept.
3. Push Back.
The applications installed on the instrument have been pre-defined to optimize the measurement
process. Changing these default application parameters will have an effect on the measurement
process and measurement results. Only qualified personnel should change these parameters or
use the instrument in expert mode.
Note: The parameters available for change will vary depending on the instrument configuration and
the application being edited.
1. From the main menu, select
2. Select Applications followed by Edit.
C A U T I O N
C A U T I O N
Settings.
Change the application settings
on page 19). When the instrument is set to off, sets
N O T I C E
Settings.
on page 19),
Change the
19