Maico PILOT TEST Operation Manual page 31

Table of Contents

Advertisement

Table 11 PILOT TEST menu
ITEM
Binaural
presentation
Show picture
Show results
Fast PILOT TEST Set On to start the test at 25 dB rather than at 50 dB.
Pause length
Print Size
Show Group 0
Output (Air
Conduction/
Free Field)
5.5.4 Audiometry menu
Figure 20
Rev.
8504660
3
Operation Manual PILOT TEST
EXPLANATION
The option to present the words to both ears at the same time.
When menu option is On, B (i.e. binaural) is a selection option
with the L/R
key. When menu option Off is selected, binaural
selection is not available from L/R
Set On/Off to show/hide picture of the test word on the display.
NOTE: When this setting is On verify the child is unable to see the
display screen.
Results table displays within the PILOT TEST screen when this
setting is On.
Pause length between the test sentences can be set between 1 s
and 20 s. Default length is 3 s.
Selection is between Short where only numerical data will print
or with Pictures where the test word picture is shown. A
checkmark ✓ is displayed if the examiner scored a correct
response. See Section 5.4.2 for more information.
When set On, an additional Group 0 is available. This is a
conditioning group where all words are presented at 50 dB.
Output selection with headphones or Free Field. Free field allows a
speaker (optional accessory) to be connected and is useful during
the conditioning of the test process. This allows the child to hear the
signal along with the examiner during training. Once testing starts,
free field should not be used for presentation.
The Audiometry menu (Figure 20) covers all
settings for performing audiometry tests. The
options are explained in Table 12.
key selection.
29
06/11/2019

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PILOT TEST and is the answer not in the manual?

Table of Contents