Automatic clear
The tare is automatically cleared after emptying the scale, if this feature is enabled. Refer to
parameter [ 113 ].
Automatic zero point correction
Zero point is corrected automatically for minor deviations if the change is within the range of limited
zeroing range value. Disable this correction at the applications like tank weighing, batching, filling
etc. against wrong zeroing at feeding. Refer to Automatic Zero Tracking parameter [ 214 ].
Automatic zero point correction range is limited together with zeroing range above.
Automatic zero at power on
Warning:
Automatic zeroing at power on can be enabled only if the scale is always unloaded
before power on. This function should be disabled at tank / silo weighing applications.
Zero point is corrected automatically at power on the instrument to compensate zero drifts of the
scale if the scale is always power on when unloaded. This feature should be disabled for tank, silo,
hopper scales etc.
Power on zero has a limited range and the instrument announces [ E E E ] error prompt in case of
out of range. Press < F > key to read the residual drift and to start indication without zeroing and
call service.
Restore Tare at Power on
This function saves the tare at power off and the instrument start to operate Net after power on as
before power off. This feature should be enabled, if taring is used at tank weighing applications.
Refer to parameter [ 112 ].
Printing
Press < Print > key when the item is on the pan and weight is stable to print the label. Refer to
Page 36 for details.
BX11 Smart Process Indicator, User Manual, Rev. 1.8, May 2017