[310 ] Zero Adjustment
This parameter is only being used for refreshing the zero level of the scale to prevent wrong
weightings from zero drifts.
1. Press < Enter > at the [ 310 ] prompt to start the zero adjustment.
2. At the [ ZEro.CA ] prompt, remove any weight on the platform, then press < Enter >.
3. The terminal automatically starts to capture zero and the [ WAıt ] message will shown
on the display 10 seconds while zero calibration is performed.
4. The following parameter is displayed after calibration. If any, error message is
announced. (Page 45)
[311 ] Span Adjustment
This parameter lets you to perform span adjustment.
1. Press < Enter > at the [ 311 ] prompt to start the span adjustment.
2. At the [ XXXXXX ] prompt, the test weight value will be used for the calibration seen
on the display. If the value of the test weights that will be used is different from the
value shown on the display, type the new value via tare and zero keys. A minimum of
20% of scale capacity is necessary for calibration; BAYKON recommends 50 to 100%.
A calibration error is produced if insufficient weight is used.
3. Place the test weights or another practical weight on the scale.
4. Press < Enter > to start span calibration. [ WAıt ] message will shown on the display
10 seconds while span calibration is performed.
5. The following parameter is displayed after calibration. If any, error message is
announced. (Page 45)
[312 ] Span Adjustment Under Load
This parameter is being used to perform span adjustment of a scale without lifting the load
on it. This operation especially used for span adjustment for filled tanks. You can make span
adjustment without emptying the tank.
1. Press < Enter > at the [ 312 ] prompt to start the span adjustment under load.
2. [ T.ZEro ] prompt appears on the display to indicate the scale load will be determined
as temporary zero.
3. Press < Enter > key and the display will show [ WAıt ] message during temporary
zero adjustment.
4. At the [ LoAd ] a little while and then [ XXXXXX ] prompt, the test weight value will
be used for the calibration seen on the display. If the value of the test weights that
will be used is different from the value shown on the display, type the new value via <
Tare > and < Zero > keys.
5. Place the test weights or another practical weight on the scale.
6. Press < Enter > to start span calibration. [ WAıt ] message will shown on the display
10 seconds while span calibration is performed.
7. The following parameter is displayed after calibration. If any, error message is
announced. (Page 45)
BX11 Smart Process Indicator, User Manual, Rev. 1.8, May 2017
Page 28 of 48