Implant Removal (Optional)
3
Remove MultiLoc screws
Instruments
03.019.005
Handle with Quick Coupling, 150 mm
03.019.010
Extraction Shaft for 4.5 mm Titanium
MultiLoc Screws, quick coupling
Assemble the extraction shaft and handle and remove all
MultiLoc screws.
Note: If 3.5 mm locking screws were used, remove them first.
4
Remove 4.0 mm locking screws
Instrument
03.019.020
Screwdriver StarDrive, T25, with coupling for
03.019.019, 330 mm
Remove all 4.0 mm locking screws with the T25 StarDrive
screwdriver.
Note: Ensure the extraction screw is connected to the nail
before removing the last locking screw.