Insert 3.5 Mm Locking Screw - Synthes MultiLoc Proximal Humeral Nail Technique Manual

Proximal humeral nail with multiple locking options for simple and fractures
Table of Contents

Advertisement

Proximal Locking
3.5 mm Locking Screws
3

Insert 3.5 mm locking screw

Instruments
03.019.005
Handle with Quick Coupling, 150 mm
314.116
StarDrive Screwdriver Shaft, T15, quick coupling
511.773
Torque Limiting Attachment,1.5 Nm,
quick coupling
Assemble the torque limiting attachment, handle and screw-
driver shaft.
Select the appropriate length 3.5 mm locking screw and
insert it through the centering sleeve. Tighten the screw until
a click is heard.
Remove the centering sleeve.
Caution: Always use a torque limiting attachment when
inserting the 3.5 mm locking screws.
Note: The centering sleeve can only be removed when the
3.5 mm locking screw is fully seated.