SETTINGS AND CALIBRATION
LEVELSCOUT INSTR
Click Measure and Aqua4Plus 2.0 will take 10 readings and display the average. Watch for
stability while the software is measuring to ensure an accurate calibration.
Click Accept and it will calculate a new offset, click Apply to accept the new offset and
Aqua4Plus 2.0 will provide a real time reading to verify calibration was successful.
Removing calibration setup values
To return to factory default simply enter an offset of 0 and slope of 1 in the Offset and
Slope boxes:
Click Apply to confirm settings and Aqua4Plus 2.0 will provide a real time reading to
confirm.
Adjusting for specific gravity
You may enter the Inverse (1/SG) specific gravity of your fluid in the Slope field to adjust
for specific gravity when needed. Click Apply to confirm settings and Aqua4Plus 2.0 will
provide a real time reading to confirm.