SETTINGS AND CALIBRATION
LEVELSCOUT INSTR
Aqua4Plus 2.0 will take 10 readings and display the average. Watch for stability while the
software is measuring to ensure an accurate calibration.
Click Ok to accept the reading and Aqua4Plus 2.0 will calculate a new pressure offset. Click
Apply to confirm the new offset value and Aqua4Plus 2.0 will provide a real time reading
to verify calibration was successful:
2 Point submergence calibration ONLY RECOMMENDED IF YOU HAVE AN ACCURATE
PRESSURE REFERENCE. Our Smart Sensors rarely change slope during normal use,
however if you have an accurate pressure source it is possible to perform a 2 point
calibration on the pressure channel.
Select 2 point Calibration under Calibration Type
Perform first point calibration as listed above and click Next