SETTINGS AND CALIBRATION
LEVELSCOUT INSTR
Enter known pressure value in matching units in the Ref. Point box for second point
measurement
Aqua4Plus 2.0 will take 10 measurements and display the average. Watch for stability
while the software is measuring to ensure an accurate calibration.
Click Ok and Aqua4Plus 2.0 will calculate the new slope and offset values. Click Apply to
confirm the new slope and offset values and the software will provide a real time reading
to verify calibration was successful.
Groundwater Elevation
Position the sensor in its desired location and ensure all data has been uploaded and
erased from the sensor before proceeding with calibration.
Once positioned connect to sensor and select the calibration button, followed by selecting
the Groundwater Elevation option. Double check measurement units selection before
proceeding.
Enter your current Groundwater Elevation reading in the Ref. Point box, making sure to
match measurement units.
Click Measure and Aqua4Plus 2.0 will take 10 readings and display the average. Watch for
stability while the software is measuring to ensure an accurate calibration.
Click Accept and Aqua4Plus 2.0 will calculate a new offset, click Apply to accept the
new offset and Aqua4Plus 2.0 will provide a real time reading to verify calibration was
successful.
Staff Gauge
Position the sensor in its desired location and ensure all data has been uploaded and
erased from the sensor before proceeding with calibration.
Once positioned connect to sensor and select the calibration button, followed by selecting
the Staff Gauge option. Double check measurement units selection before proceeding.
Enter your current Staff Gauge reading in the Ref. Point box, making sure to match
measurement units.