Battery Warnings; Battery Charger Warnings - Integra 90520 Operation And Service Manual

Led surgical headlight system
Table of Contents

Advertisement

Battery Warnings

• Defective batteries can damage the Surgical Headlight
System. If the battery shows any signs of damage, leakage,
or cracking, it must be replaced immediately and only with
an Integra battery recommended for or supplied with
the system.
• Improper disposal of batteries may create an explosion or
contamination hazard. Never dispose of batteries in refuse
containers. Do not dispose of the battery in fire. Always
recycle batteries according to local regulations.
• Improper handling of the battery can lead to heat
generation, smoke, bursting, or fire.
• Do not remove the label from the battery.
• Do not immerse the battery in water or other liquids.
• Do not disassemble, modify, or solder the battery.
• Do not directly connect or short-circuit the positive (+) and
negative (-) battery terminals.
• To avoid short circuits, keep battery terminals away from
metal objects.
• Do not expose the battery to temperatures higher than
49C/120F.
• You must charge the batteries for four (4) hours or until
completely charged, before using the battery power option.
• A dead battery may result if the batteries are left uncharged
or unused for a long period of time.
• If the LED Surgical Headlight System will be unused for
several months or longer, remove the battery from the
battery holster prior to storing the system.

Battery Charger Warnings

• Do not expose the charger or power supply to water or
liquids. The case is not sealed.
• Do not open the charger or power supply case. No user
serviceable parts are inside.
• Do not cover the fan exhaust or obstruct the airflow, this will
cause overheating.
• Use only the manufacturer's recommended power supply,
charger, batteries and power cords.
• Store the charger in a cool spot, away from external heat
sources. If replacement is necessary, replace only with
approved power cord.
Caution
• During recharging, the charger may become warm.
• No modification of this equipment is allowed.
• This device should be used in environments that meet the
operating environmental conditions in the specification
section of this manual.
6