Type
Wattage
Lamp Life
Color Temperature
Spot Size Range
Power Input
Battery Power Kit
Type
Capacity
Voltage
Electrical Safety
Battery Charger, Single Bay and Dual Bay
Classifications
Input
Output
Direct Power Kit
Classifications
Power Input
Power Output
AC Power Leakage
Electromagnetic Compatibility
Power Cords
Type
Appliance End
Outlet End
Environment
Operating Temperature
Operating Humidity
Storage/Transport Temperature
Storage/Transport Humidity
Mode of Operation
Ingress Protection
Pressure
15
LED
13 Watts
>25,000 hours
Average 4500 (minimum of 4300)
25mm to 125mm at 16 in (~40cm) working distance
9-15 VDC
Lithium Ion, 4 cell pack
37-45 Wh
14.4 VDC @ 2.6-3.1 Ah
IEC 60601-1
US/FDA, Canada, Australia, EU: Class II, Double Insulated
100-240 VAC, 47-63 Hz, 1.5 A
24 VDC, 2.7 A
US/FDA, Canada, Australia, EU: Class II, Double Insulated
100-240 VAC, 47-63 Hz, 1.0 A
15 VDC, 2.0 A
<100μA @ 264VAC, 60Hz
EN55022/55024/61000
Class II, 2.5-3.0A, Double Insulated, Detachable
IEC 60320 C17
Country specific for US, UK, AU, EU
6 C to 40 C (43 F to 104 F). Above 5,000 ft (1,524 m) elevation, the
maximum operating temperature is 32 C (90 F).
10% to 95% Relative Humidity, non-condensing
-20 C to 49 C (-4 F to 120 F)
30% to 80% Relative Humidity, non-condensing
Continuous
IP20
Operating 700-1060hPa / Storage 500-1060hPa