Media compatibility
316 SS
Temperature limits
Operating temperature
–10 °C to 50 °C (14 °F to 122 °F)
Up to 95 % RH non-condensing.
No change in accuracy over operating temperature range.
Gauge must be zeroed to achieve rated specification.
Process temperature
–10 °C to 50 °C (14 °F to 122 °F)
Warm-up time
Five (5) minutes.
Accuracy Statement
Absolute sensor
0 % to 110 % of Range:
0 % to 110 % of Range:
Compound sensor
0 % to 110 % of Range:
0 % to 110 % of Range:
Vacuum*:
*Vacuum = –14.5 psi
User Manual for meriGauge Plus System
± (0.02 % of Full Scale) +
(0.005 % of Reading)
± (0.02 % of Full Scale) +
(0.005 % of Reading)
± (0.02 % of Full Scale)
15 of 47