Measure S-Parameters (NA Mode)
S-parameter syntax is described by the following:
S (out | in)
out = FieldFox receiver port
in = FieldFox source port
1. Press Preset then
2. Press Measure 1 then choose from the following:
1-port reflection measurement.
S11
2-port transmission measurement.
S21
Reverse 2-port transmission measurement.
S12
Requires Opt 122.
Reverse 1-port reflection measurement.
S22
Requires Opt 122.
OR Press Trace 6 then
Multi-trace configuration. Learn more on page 8.
3. Press Freq/Dist then either
and
Center
measurement.
4. Press BW 2 then
for the measurement. Narrower bandwidths require
more time to sweep.
5. Press Sweep 3 then
number of data points for the measurement. More
data points requires more time to sweep.
6. Press Cal 5 to calibrate the measurement. See page
6 to learn more.
Continued on following page...
Preset
Num of Traces
to enter a frequency range for the
Span
IFBW
Resolution
and
Start
to select the IF Bandwidth
to select the
6
then select a
or
Stop