Mounting; General Instructions - Vega VEGASWING 51 Operating Instructions Manual

Contactless electronic switch
Table of Contents

Advertisement

Switching point
VEGASWING 51 - - contactless electronic switch

4 Mounting

4.1 General instructions

In general, VEGASWING 51 can be mounted in any position.
The instrument must be mounted in such a way that the tuning
fork is at the height of the requested switching point.
The tuning fork has lateral marking (notches), marking the
switching point with vertical mounting. The switching point
refers to water (1 g/cm³/0.036 lbs/in³). When mounting
VEGASWING 51, make sure that this marking is at the height
of the requested switching point. Keep in mind that the
switching point of the instrument is shifted if the medium has a
density other than water.
2
1
3
Fig. 2: Vertical mounting
1
Switching point approx. 13 mm (0.51 in)
2
Switching point with lower density
3
Switching point with higher density
Mounting
11

Advertisement

Table of Contents
loading

Table of Contents