Mounting
16
neutral bending axis
neutral bending axis
initial state (unloaded)
Alternatively, thin steel plates (approx. 0.5 mm thick) can
be placed under one side, under the load introduction
parts, at the given points.
Structurally, this effect can be counteracted by shifting
the bearing points on the weighbridge as far upward as
possible, towards the neutral bending axis.
A1461-2.0 HBM: public
unloaded
loaded
Steel plates