HBM C16i Series Mounting Instructions page 12

Hide thumbs Also See for C16i Series:
Table of Contents

Advertisement

Available languages

Available languages

Mounting
10
With the C16i..., we recommend the use of EPO3/50t
and C16/EPU44A or C16/ZOU44A mounting accessories
from HBM, because they permit easy mounting. The
rotation stop device welded onto the load cells and the
dowel pin provided are also suitable for this model type
(see chapter 8 "Dimensions").
Preparation work for mounting
When using EPO3/50t, C16/EPU44A or C16/ZOU44A to
introduce and carry off the load, the following
preparations have to be made: A dowel pin is enclosed in
the packing for every load cell. This dowel pin and the
rotation stop welded onto the load cell prevent a potential
transducer microrotation, thus preventing the cable from
being damaged. Use a hammer to drive in the dowel pin
until it rests in the load introduction part's pocket bore.
Radially position the open side of the dowel pin.
CAUTION
Position of rotation stop dowel pin:
Radially position open side!
The bore depth ensures correct fitting of the pin. Only
one load introduction part per load cell must be equipped
with this pin. This load introduction part must be mounted
below the load cell to enable the dowel pin to engage into
the rotation stop recess (see chapter 8 "Dimensions"). No
A1461-2.0 HBM: public