E-664 LVPZT Position Servo Controller
8.
Rear Panel Elements
Line Switch
Voltage Selector
Fuses
Fuse carrier
Fig. 6. Fuse access
Stage Connector
Control Input
Sensor Monitor
I/O Connector
Release 1.0.4
Power On/Off (opens completely)
Two default ranges (220-240 V or 110-120 V). Selected
value is visible in window (see figures). Pry out and re-orient
fuse carrier to change range selection. Fuses will also need
to be replaced.
Fuses are voltage dependent: use 0.8 A for 230 V and 1.6 A
for 110 V (slow type)
Door with
window
Line cord
connector
Single connector for combined piezo actuator and sensor
signals
Control voltage for open / closed-loop operation
Current sensor value
Combines control input, sensor monitor, overflow and on
target signals for all three channels for easy access to
complete functionality. If this connector is used, do not
connect BNC-inputs.
Fuse access
www.pi.ws
User Manual PZ 99E
Page 17