Applying the Blood Sample

Step 8
Obtain a blood sample using the lancing
device. Place the device against the
pad of the finger. The best puncture
sites are on the middle or ring fingers.
Press the release button. Remove the
lancing device from the finger. Wait
a few seconds for a blood drop to form.
A minimum volume of 0.5 microliter is needed to fill the
confirmation window.
Step 9
After the
symbol appears on
the screen, apply the blood sample to
the narrow end of the test strip until
the meter beeps. If the blood sample
does not fill the confirmation window
completely, an Er4 message may appear
because of abnormal viscosity or
insufficient volume.
Then discard the test strip, insert a new
one, and apply the blood sample again
after
symbol appears on the
screen.
26 www.i-sens.com
Approximate size
Good
Insufficient
Sample
Sample
Note:
The meter will switch off if the blood sample is not
applied within 2 minutes of the
the screen. If the meter turns off, press S button or button
and apply the blood sample after the
on the screen.
Step 10
The test result will appear after the meter
counts down from 5 to 1. The result will be
automatically stored in the meter memory.
If the test strip is removed after the test
result has been displayed, the meter
will automatically switch off after 3
seconds. Discard used test strip safely in a
disposable container.