Checking The System; Control Solution Testing - i-SENS CareSens N User Manual

Blood glucose monitoring system
Hide thumbs Also See for CareSens N:
Table of Contents

Advertisement

Available languages

Available languages

Checking the System

You should check your meter and test strips using
CareSens Control Solution (Control A and B). The
CareSens Control Solutions contain known amounts
of glucose and are used to check that the meter and
the test strips are working properly. The test strip
Control Solution
vials have CareSens Control Solution ranges printed
on their labels. Compare the result displayed on
EXP
the meter to the control solution range printed on
Control Solution Range Written on Strip Vial.
Discard Date :
Manufacturer : i-SENS, Inc.
43, Banpo-daero 28-gil, Seocho-gu, Seoul, 06646, Korea
the test strip vial. Before using a new meter or a
PGA2E0580 REV8 09/2015
new vial of test strips, you should conduct a control
solution test following the procedure on page 18 with two
different levels of solutions (Control A and B).
Notes:
• Use CareSens Control Solutions only.
• Check the expiration dates printed on the bottle. When you
first open a control solution bottle, record the discard date
(date opened plus three (3) months) in the space provided on
the label.
• Make sure your meter, test strips, and control solutions are at
room temperature before testing. Control solution tests must
be done at room temperature (68 - 77˚F, 20 - 25˚C).
• Before using the control solution, shake the bottle, discard the
first 1 or 2 drops and wipe the tip clean.
• Close the control solution bottle tightly and store at a
temperature of 46 - 86˚F (8 - 30˚C).
You should do a control solution test:
• When you want to practice the test procedure using the
control solution instead of blood;
• When using the meter for the first time;
18 www.i-sens.com
• Whenever you open a new vial of test strips;
• If the meter or test strips do not function properly;
• If your symptoms are inconsistent with the blood glucose
test results and you feel that the meter or test strips are not
working properly;
• If you drop or damage the meter.

Control Solution Testing

Step 1
Insert a test strip into the meter's test strip
port, with the contact bars facing upwards.
Gently push the test strip into the port until the
meter automatically turns on and the
symbol appears.
Step 2
Shake the CareSens Control
Solution bottle well before each
test. Remove the cap and squeeze
the bottle to discard the first 1 or 2
drops.
Then wipe the tip with a clean tissue
or cloth. While the
symbol
blinks, apply the control solution to
the narrow edge of the test strip until the meter beeps. Make
sure the confirmation window of test strip fills completely.
www.i-sens.com 19