GE MEDICAL SYSTEMS
REV 10
4–3–3 Utility Menu (Continued)
(C) TROUBLE IMAGE SAVE/LOAD/DISPLAY (Continued)
Procedure for Saving Trouble Images
1. Type 1 and then press Return key. The message "Set a media to drive and then press any key." is displayed.
2. Set a floppy disk into the FD drive and press any key in the case of system software version 1.20 or former
console.
Set an 128MB MO (Magneto Optical) disk into the MOD drive and press any key in the case of system software
version 2.00 or later console.
The Trouble Image File Menu will be displayed. Refer to ILLUSTRATION 4–15.
The 230MB MO disk is available on the console with software version 3.10 or later.
3. Input the menu number you desire to save into the disk.
4. Press Return key.
During the system is saving the file, "Saving now" is displayed on the CRT monitor. After saving the file,
"Succeeded" or "Failed" is displayed on the CRT monitor. Then, "Press Y to continue, Control + R to quit" is dis-
played. Type Y and press Enter to continue to use this function. Press and hold Control, and press R to quit this
function.
Procedure for Loading Trouble Image
1. Type 2 and then press Return key. The message "Set a media to drive and then press any key." is displayed.
2. Set the floppy disk storing the previous trouble images into the FD drive and press any key in the case of sys-
tem software version 1.20 or former console.
Set the 128MB MO (Magneto Optical) disk storing the previous trouble images into the MOD drive and press
any key in the case of system software version 2.00 or later console.
The Trouble Image File Menu will be displayed. Refer to ILLUSTRATION 4–15.
The 230MB MO disk is available on the console with software version 3.10 or later.
3. Input the menu number you desire to load from the floppy or MO disk.
4. Press Return key.
During the system is loading the file, "Loading now" is displayed on the CRT monitor. After loading the file,
"Succeeded" or "Failed" is displayed on the CRT monitor. Then, "Press Y to continue, Control + R to quit" is dis-