GE MEDICAL SYSTEMS
REV 14
6–2–11 CRT Cable Set (FRU No. 117)
Time Required
0.5 Hour
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6–23.
1.
Turn OFF the system.
2.
Remove the Monitor Assy (FRU 100). Refer to 6–2–1 on page 6–66.
3.
Remove the Monitor Cover Set (FRU 101). Refer to 6–2–2 on page 6–68.
Disconnect three connectors ( 1 – 3 ) from the back side of the CRT Assy.
4.
CRT ASSY DISASSEMBLY
ILLUSTRATION 6–23
6–82
LOGIQ 500 SERVICE MANUAL
P9030TA
RENEWAL PARTS