Troubleshooting

If you follow the recommended action but the problem persists,
or error messages other than the ones below appear, please call
your local customer service.

Result Readings

Message
< 10 mg/dL (0.5 mmol/L)
≥ 240 mg/dL (13.3 mmol/L)
> 600 mg/dL (33.3mmol/L)

Error Message

Error Message Cause
The batteries cannot
provide enough power
for a test.
What it Means
What To Do
Replace the batteries
immediately.
Strip has been used.
Problem in operation.
You may have removed
the strip after applying
blood, or insufficient
blood volume.
HCT>75
Wrong code strip
is inserted or other
coding errors.
Expired code strip.
EN-11
ADVANCED pro GD40g Owner's Manual
Repeat the test with a
new strip.
Review the instructions
and repeat the test with
a new strip. If problem
persists, contact the local
customer service for
assistance.
Review the instructions
and repeat test with a
new test strip.
Make sure to insert the
right code strip for the
meter. Check if the code
numbers on the code strip
and the strip vial label are
the same.
Repeat the test with a
new code strip.