ADVANCED pro GD40g Owner's Manual
• Keep the strip vial away from children since the cap and the test
strip may be a choking hazard. If swallowed, promptly see a doctor
for help.
For further information, please refer to the test strip package insert.
Important Control Solution Information
• Use only our control solutions with your device.
• Do not use the control solution beyond the expiry date or 3 months
after first opening. Write the opening date on the control solution
vial and discard the remaining solution after 3 months.
• It is recommended that the control solution test be done at room
temperature 20°C to 25°C (68°F to 77°F). Make sure your control
solution, device, and test strips are at this specified temperature
range before testing.
• Shake the vial before use, discard the first drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
• Store the control solution tightly closed at temperatures between
2°C to 30°C (35.6°F to 86°F). Do NOT freeze.
Reference Values
The device provides you with plasma equivalent results.
Time of day
Fasting and before meal
2 hours after meals
Source: American Diabetes Association (2012). Clinical Practice
Recommendations. Diabetes Care, 35 (Supplement 1): S1-100.
Normal plasma glucose range for
people without diabetes (mg/dL)
< 100 mg/dL (5.6 mmol/L)
< 140 mg/dL (7.8 mmol/L)
Normal results vary of Hematocrit, but in general are as follows:
• Male: 40.7 to 50.3%
• Female: 36.1 to 44.3%
If the test result of hematocrit is more than 70%, or hemoglobin is
more than 23.8 (g/dL), the blood glucose test may be invalid, please
redo the test, and consult your doctor if the result is repeatedly
invalid.
Please consult your doctor to determine a target range that
works best for you.
Symbol Information
Symbol Referent
For in vitro diagnostic
use
Do not reuse
Consult instructions
for use
Keep away from
sunlight
Keep dry
Temperature limitation
Use by
EN-10
Symbol Referent
Use within 6 months
after first opening
6M
Batch code
Manufacturer
Serial number
Caution, consult
accompanying
documents
Sterilized using
irradiation
Do not use if package
is damaged