Before Testing

Calibration
You must calibrate the deviced every time you begin to use a new
vial of test strips by setting the meter with the correct code. To
ensure test accuracy, make sure the code number displayed on
the display screen matches the number printed on the strip vial.

How to Code Your Meter

1. Insert the code strip into the test slot of the device. Wait for the
device to display the code number.
Important!
• Make sure the code number on display, code strip, and test strip
(c)
vial are the same.
2. Remove the code strip, the display will show "OFF" indicating
the device has finished coding and is ready for blood glucose
testing.

Checking the Code Number

Make sure that the code number displayed on the device matches
the number on the test strip vial before you proceed. (d)
If it matches, you can proceed with your test. Otherwise, please
stop testing and repeat the calibration procedure.
If the problem persists, contact Customer Service for further
assistance.
Important!
• It is important to make sure that the LCD displayed code is the same
as the code on the test strip vial before testing. Failure to do so will
get inaccurate results.

Control Solution Testing

Our Control Solution contains a known amount of glucose that
reacts with test strips and is used to ensure your device and test
strips are working together correctly.
Do a control solution test when:
you first receive the device.
9
9
at least once a week to routinely check the device and test strips.
9
9
you begin using a new vial of test strips.
9
9
you suspect the device or test strips are not working properly.
9
9
your blood glucose test results are not consistent with how
9
9
you feel, or if you think the results are not accurate.
you have dropped or think you may have damaged the device.
9
9
To perform the control solution test, do the following:
1. Insert the test strip into the test slot of the device. Wait for the
device to display the test strip and blood drop symbol. (e)
2. Press MAIN to mark this test as a control solution test.
With "QC" displayed, the device will store your test result in
memory. If you press again, the "QC" will disappear and this test
is no longer a control solution test. (f)
Important!
• When doing the control solution test, you have to mark it so that
the test result will not be stored in the memory. Otherwise, it will
mix up the blood glucose test results with the control solution test
results in memory.