Adjusting The Aperture Diaphragm (As); Attaching The Darkfield Ring Slit (For Darkfield Observation Only) - Olympus CX33 Instructions Manual

Biological microscope
Table of Contents

Advertisement

8

Adjusting the aperture diaphragm (AS)

a
9

Attaching the darkfield ring slit (for darkfield observation only)

c
2
b
Using the eye shades
When wearing eyeglasses
Use the eye shades in the normal, folded-down position.
When not wearing eyeglasses
Extend the folded eye shades in the arrow direction. Since the eye
shades prevent the unnecessary light from entering between eyepieces
and eyes, you can observe the specimen comfortably.
The aperture diaphragm is a diaphragm to adjust the numerical aperture
of the condenser.
Adjusting the numerical aperture of the condenser appropriately
with respect to the numerical aperture of each objective allows
you to observe the specimen with the best contrast and resolution.
(For the aperture diaphragm, see page 22.)
1
NOTE
For darkfield observation, the darkfield ring slit CH2-DS (option) and the filter
holder CH2-FH (option) must be attached to the condenser.
1
a
2
1
NOTE
Move the aperture diaphragm lever a to right and left direction to
match the index with the magnification of the objective in use.
For darkfield observation, open the aperture diaphragm.
Insert the darkfield ring CH2-DS a into the filter holder CH2-FH b .
Push the filter holder b which includes the darkfield ring into the bottom
of the condenser c until the click sound is heard.
The darkfield observation is not available with the 100X oil
immersion objective.
CX33
17

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents