Emerson ROSEMOUNT 5081-A Instruction Manual page 56

Fieldbus two-wire chlorine, dissolved oxygen, and ozone transmitter
Table of Contents

Advertisement

MODEL 5081-A
7.5.4 Procedure using Deltav
Access: deltaV Explorer/transducer Block/Properties, Amperometric Sensor tab
1. Parameter: Amperometric Stabilize time (AMP_SPAN_StABILIZE_tIME)
Set the stabilization time between 0 and 99 seconds. the default value is 10 seconds.
2. Parameter: Amperometric Stabilize Value (AMP_SPAN_StABILIZE_VALuE)
Set the stabilization range to between 0.01 and 9.99 ppm. the default values are shown in the table:
oxygen
0.05 ppm or 1%
Free chlorine
0.05 ppm
total chlorine
0.05 ppm
ozone
0.01 ppm
3. Parameter: Salinity (SALINItY).
Enter the salinity of the water.
4. Parameter: Zero Limit (AMP_ZERo_StABILIZE_VALuE)
Enter the desired zero limit. default limits are given in the table:
oxygen (ppm)
oxygen (ppb)
oxygen (% saturation)
Free chlorine
total chlorine
ozone
50
0.05 ppm
2.0 ppb
1%
0.05 ppm
0.05 ppm
0.01 ppm or 10 ppb
SECTION 7.0
PROGRAMMING