7.3.3 Procedure using the infrared remote controller
1. Press PRoG on the remote controller.
2. Press NEXt until the tEMP submenu appears. Press ENtER.
3. the screen displays the tAUtO (automatic temperature compensation) prompt.
Press é or ê to toggle between On and OFF. Press ENtER to save.
4. If you disable tAuto, the tMAN prompt appears. use the arrow keys to change the
temperature to the desired value. to enter a negative number, press è or ç until no
digit is flashing. then press é or ê to display the negative sign. The temperature
entered in this step will be used in all measurements (oxygen, chlorine, ozone,
or pH), no matter what the process temperature is. Press ENtER to save.
5. Press RESEt to return to the process display.
7.3.4 Procedure using Deltav
Access: deltaV Explorer/transducer Block/Properties/temperature Compensation
1. Parameter: Auto/Manual Selection (SENSoR_tEMP_CoMP)
Select automatic or manual temperature compensation
2. If manual temperature compensation was chosen:
Parameter: Manual temperature (SENSoR_tEMP_MAN_VALuE)
Enter temperature value to be used with manual temperature compensation.