Emerson ROSEMOUNT 5081-A Instruction Manual page 49

Fieldbus two-wire chlorine, dissolved oxygen, and ozone transmitter
Table of Contents

Advertisement

MODEL 5081-A
ITEM
A. Temperature compensation
1. Automatic
2. Manual
B. Display
1. type of measurement
2. units (oxygen only)
3. units (ozone only)
4. Sensor (oxygen only)
5. temperature units
6. output current units
7. Security code
C. Calibration Setup
1. Stabilization criteria
a. time
b. change
if oxygen (ppm or ppb)
if oxygen (%)
if ozone
if chlorine
2. Salinity (oxygen only)
3. Slope (chlorine only)
4. Maximum zero limit
a. if oxygen (ppm)
b. if oxygen (ppb)
c. if oxygen (%)
d. if ozone
e. if chlorine
D. Line Frequency
E. HART
1. Address
2. Preamble
3. Burst
4. Id
F. pH Settings (free chlorine only)
1. Automatic pH correction
2. Manual pH correction
3. Location of preamplifier
4. pH sensor diagnostics
a. reference offset
b. diagnostics
(1) glass impedance temperature
correction
(2) glass impedance high
(3) glass impedance low
5. Calibration settings
a. automatic buffer calibration
b. buffer selection list
c. stabilization criteria
(1) time
(2) change
G. Pressure settings (oxygen only)
a. units
b. pressure for % sat calculations
TABLE 7-1. Default Settings
MNEMONIC
tEMP
tAUtO
tMAn
dISPLAy
tyPE
oxygen, ozone, free chlorine, total chlorine
Unit
Unit
SEnSor
499Ado, 499Atrdo, Hx438 or Gx338, other biopharm
tEMP
OutPut
CodE
CAL SEtUP
StAbiLiSE
tiME
dELtA
SALnty
SLOPE
LiMit
LinE FrEq
HArt
AddrESS
PrEAMb
bUrSt
Id
pH
MAn
PAMP
dIAgnOStIC
rOFFSEt
diAG
IMPtC
GFH
GFL
PH CAL
bAUtO
buFFEr
StAbiLiSE
tiME
dELtA
BAr PrESS
Unit
% SAt P
CHOICES
on or off
-25.0 to 150°C
ppm, ppb, or %
ppm or ppb
°C or °F
mA or % of full scale
000 to 999
00 to 99 sec
0.01 to 9.99 ppm
1 to 100 %
0.01 to 9.99 ppm
0.01 to 9.99 ppm
0.0 to 36.0
single or dual
00.00 to 10.00 ppm
000.0 to 999.9 ppb
000.0 to 999.9 %
00.00 to 10.00 ppm
00.00 to 10.00 ppm
50 or 60 Hz
00 to 15
05 to 20
on or off
0000000 to 9999999
on or off
0.00 to 14.00
transmitter or sensor
0 to 999
on or off
on or off
0 to 2000 MW
0 to 900 MW
on or off
see table in Section 13.1
0 to 99 sec
0.02 to 0.50
mm hg, kPa, atm, bar, in Hg
0 to 9999
SECTION 7.0
PROGRAMMING
DEFAULT
on
25°C
oxygen
ppm
ppm
499Ado
°C
mA
000
10 sec
0.05 ppm
1 %
0.01 ppm
0.05 ppm
0.0
single
0.05 ppm
2.0 ppb
1%
0.01 ppm
0.05 ppm
60 Hz
00
05
off
0000000
on
7.00
transmitter
60
off
on
1000 MW
10 MW
standard
10 sec
0.02
mm Hg
760 mm Hg
43