Emerson ROSEMOUNT 5081-A Instruction Manual page 53

Fieldbus two-wire chlorine, dissolved oxygen, and ozone transmitter
Table of Contents

Advertisement

MODEL 5081-A
PRoGRAM
Unit
PPb
EXIt
ENtER
7.4.4 Procedure using Deltav
1. Access: deltaV Explorer/Context menu
Change Process Variable type (method_change_pv_type).
Select desired measurement (PV)
Although monochloramine is a choice, a monochloramine sensor is NOT currently available from Rosemount
Analytical.
2. If you chose o
, select a sensor from the following table:
2
AdO
499Ado
trdO
499Atrdo
SdO1 Hx438 or Gx448 steam-sterilizable sensor
SdO2 Steam-sterilizable sensor from other manufacturer
Access: deltaV Explorer/transducer Block/Properties, Amperometric Sensor tab
Parameter: oxygen Sensor type (SENSoR_tYPE_oXYGEN)
Select desired sensor
3. on the Amperometric Sensor tab -
Parameter: Primary Value unit (PRIMARY_VALuE_uNIt)
Select the desired units. For best results, make the following settings based on the sensor used:
Sensor
Units
499Ado
ppm or %
499Atrdo
ppb
Gx448
ppm or %
Hx438
ppm or %
6. For best results make the following settings based on the sensor being used.
Sensor
Units
499Ado
ppm or %
499Atrdo
ppb
Gx448
ppm or %
Hx438
ppm or %
7. If you chose O3 in step 3, the screen at left appears. Press é or ê to toggle between
ppm and ppb. Press ENtER to save.
8. Press RESEt to return to the main display.
SECTION 7.0
PROGRAMMING
47