Emerson ROSEMOUNT 5081-A Instruction Manual page 104

Fieldbus two-wire chlorine, dissolved oxygen, and ozone transmitter
Table of Contents

Advertisement

MODEL 5081-A
15.2 TROUBLESHOOTING WHEN A FAULT, WARNING, OR ERROR MESSAGE IS SHOWING
Fault
Explanation
RTD OPEn
Rtd measuring circuit is open.
bAd rtd
Rtd resistance is outside the range expected for a Pt100 or 22k NtC.
PHgLASS HI
pH glass impedance exceeds programmed limit.
PHgLASS LO
pH glass impedance is below programmed limit.
AdC FAIL
Analog to digital conversion has failed.
Warning
Explanation
OuEr rAngE
Process variable exceeds the display limit.
In Curr HI
Sensor input current exceeds 210 uA.
In Curr LO
Sensor input current is a large negative number.
NEED 0 CAL
Sensor needs zeroing. Concentration is a large negative number.
tEMP HI
temperature reading exceeds 150°C.
tEMP LO
temperature reading is less than -15°C
SEnSE OPEn
Rtd sense line is open or not connected.
PH In HI
mV signal from pH sensor is too big (chlorine only).
NO SOLngnd
Solution ground terminal is not connected.
EECHECSUn
An EEPRoM byte changed unexpectedly.
EE OF
during setup or burn, EEPRoM command list overflowed.
EE Error
EEPRoM byte failed to verify.
bAd Gnd
Bad ground exists.
FACtCAL
transmitter needs factory calibration.
Error
Explanation
SLOPE HI
Glass electrode slope exceeds 62 mV/pH.
SLOPE LO
Glass electrode slope is less than 40 mV/pH.
0 OFFSEt
Zero offset exceeds programmed limit.
CAL ErrOr
Amperometric sensor sensitivity (nA/ppm) is very large or very small
EEProtECt
EEPRoM is write protected
98
SECTION 15.0
TROUBLESHOOTING
See Section
15.2.1
15.2.1
15.2.2
15.2.2
15.2.3
See Section
15.2.4
15.2.4
15.2.4
15.2.5
15.2.1
15.2.1
15.2.1
15.2.6
15.2.7
15.2.8
15.2.9
15.2.10
15.2.11
15.2.12
See Section
15.2.13
15.2.13
15.2.14
15.2.15
15.2.16