Programming; General; Default Settings; Temperature Settings - Emerson ROSEMOUNT 5081-A Instruction Manual

Fieldbus two-wire chlorine, dissolved oxygen, and ozone transmitter
Table of Contents

Advertisement

MODEL 5081-A

7.1 GENERAL

this section describes how to do the following:
1. enable and disable automatic temperature correction
2. program the type measurement (oxygen, ozone, or chlorine)
3. setup stabilization criteria for calibration
4. enable automatic pH correction for chlorine measurements
5. choose units for barometric pressure (oxygen only)
6. choose limits for diagnostic fault messages
Each section contains definitions of terms used, programming instructions using the infrared remote controller, and pro-
gramming instructions using deltaV.

7.2 DEFAULT SETTINGS

table 7-1 lists the default settings for the 5081-A transmitter. the transmitter is configured at the factory to measure
oxygen.
Before changing any default settings, configure the transmitter for the measurement you want to
make: oxygen, free chlorine, total chlorine, or ozone. Changing the measurement ALWAyS returns
the transmitter to factory default settings.
42
SECTION 7.0

PROGRAMMING

7.1
General
7.2

Default Settings

7.3

Temperature Settings

7.4
Display
7.5
Calibration Setup
7.6
Line Frequency
7.7
pH Measurement
7.8
Barometric Pressure
IMPORTANT
SECTION 7.0
PROGRAMMING