Thermo Scientific Accela Hardware Manual page 31

Ri detector
Hide thumbs Also See for Accela:
Table of Contents

Advertisement

Thermo Scientific
Figure 16. Solvent line connections for a single detector configuration
Accela
RI Detector
Accela
autosampler
Drain
manifold
Accela
pump
Convoluted
tubing
To waste
container
A
B
D
J2
J4
J3
A
B
C
D
B
C
D
Connecting the Inlet and Outlet Tubing
C
0.005 in. ID × 1/16 in. OD
stainless steel inlet tubing
In port
(connection from
LC column)
In
Out
Out port
(connection to waste)
3/32 in. OD × 0.060 in. ID
Teflon outlet tubing
Accela RI Detector Hardware Manual