Accela RI Detector Hardware Manual
Table 2. Accela RI Detector specifications (Sheet 2 of 2)
Item
Fuse size
Operating humidity
Method of detection
Refractive index range
Attenuator range
Linearity range
Noise
Rise time
Polarity
Zero adjustment
Automatic zeroing range
Resolution of autozeroing
Offset adjustable range
Resolution of offset
Integrator output
Recorder output
External communication
Event marker input
Signal output
Signal input
Temperature control
Flowcell volume
Maximum flow rate
Pressure rating
Liquid-contact materials
Specification
• For 110/115 V ac: T3.15 A, 5×20 mm
• For 230 V ac: T1.6A, 5×20 mm
5% to 95% non-condensing relative humidity
Deflection
1.00 to 1.75 RIU
0.25 to 512 μRIU
600 μRIU
–9
= 2.5 × 10
RIU (Response:1.5 seconds)
0.1, 0.25, 0.5, 1, 2. 3. 6 seconds
Positive/Negative
Optical zeroing, automatic zeroing
–5
± 10 × 10
RIU
–9
± 10 × 10
RIU
0 to 500 mV (same with Integrator output sensitivity)
10 mV (same with Integrator output sensitivity)
0 to 1V/FS (Sensitivity: 512 × 10
0 to 10 mV/FS
Ethernet
Contact: = DC 5V, 0.1 A
• Purge on/off
• Automatic zeroing on/off
• Run
Contact: = 5V, 0.1 A Ready
Run, Autozero, Purge
30 to 50 °C (1 °C increment) or off
8 μL
10 mL/min (mobile phase: deionized water)
517 kPa (75 psi)
316 stainless steel, Teflon™, Quartz optical glass
–6
RIU/FS)