Analyser Problem Solving - Kane 455 Manual

Flue gas analyser with direct co2 measurement
Hide thumbs Also See for 455:
Table of Contents

Advertisement

8.

ANALYSER PROBLEM SOLVING

If any problems are not solved with these solutions, contact us or an authorized repair
center.
Fault symptom
• Oxygen too high
• CO
too low
2
• CO reading (- - - -)
• Batteries not holding charge
• Analyser not running on mains
adapter.
• Analyser does not respond to flue
gas
• Net temperature or Efficiency
calculation incorrect.
• Flue temperature readings erratic
• T flue or T nett displays
• Ratio, EFF, X-Air all display (- - - -)
For latest prices and delivery to your door visit MyTub Ltd - www.mytub.co.uk - info@mytub.co.uk
For latest prices and delivery to your door visit MyTub Ltd - www.mytub.co.uk - info@mytub.co.uk
Causes / Solutions
• Air leaking into probe, tubing, water trap,
connectors or internal to analyser.
• Analyser was stored in a cold environment
and is not at normal working temperature.
• CO sensor needs replacing.
• Batteries exhausted.
• AC charger not giving correct output.
• Fuse blown in charger plug.
• Particle filter blocked.
• Probe or tubing blocked.
• Pump not working or damaged with
contaminants.
• Ambient temperature set wrong during
Automatic Calibration.
• Temperature plug reversed in socket.
• Faulty connection or break in cable or
plug.
• Probe not connected.
(-OC-)
• Faulty connection or break in cable or
plug.
• CO
reading is below 2%.
2
KANE455 manual
Page 27