Chapter 4 Operation

4.2.4.6 Menu inact. T/O

The Menu inactivity timeout sets the amount of time before any menus are collapsed and the default
graphic screen is displayed. Values are 1 (default) to 255 minutes. A value of zero (0) will disable this
feature.
To change the timeout value:
From the Hidden menu use the UP (↑ ↑ ↑ ↑ ↑ ) or DOWN (↓ ↓ ↓ ↓ ↓ ) button to highlight Menu inact. T/O and then press
the ENTER button. Use the numeric keypad to change the value and then press the ENTER button.

4.2.4.7 File maintenance

This option allows the user to erase or write-protect files.
To perform File maintenance:
From the Hidden menu use the UP (↑ ↑ ↑ ↑ ↑ ) or DOWN (↓ ↓ ↓ ↓ ↓ ) button to highlight File maintenance and then press
the ENTER button. Use the UP (↑ ↑ ↑ ↑ ↑ ) or DOWN (↓ ↓ ↓ ↓ ↓ ) button to select a file from the File Directory and then
press the ERASE button to erase the file or the PROT button to write-protect the file. An asterisk (*) after
the file size indicates that a file is write-protected. Press the EXIT button to return to the Hidden menu.

4.2.4.8 Flush interval

This option sets the amount of time between flushes to disk. Values are 1 to 255 minutes. Default is 15
minutes. The higher the value, the less wear there will be on the media; however, there is an exposure in
that a power failure may cause the loss of data in the buffer since the last flush to media.
To change the Flush interval value:
From the Hidden menu use the UP (↑ ↑ ↑ ↑ ↑ ) or DOWN (↓ ↓ ↓ ↓ ↓ ) button to highlight Flush interval and then press the
ENTER button. Use the numeric keypad to change the value and then press the ENTER button.
Page 4-11