Chapter 2 Installation and Wiring

2.2.1 Panel Mounting

The Recorder should be mounted in a vertical panel to ensure proper operation. Ensure you have the proper
clearances and proceed as follows:
2.2.1.1 Cut a panel opening 5.43" x 5.43" [138 mm x 138 mm].
2.2.1.2 Remove any packaging material from the Recorder. Always handle the unit carefully to avoid
damaging the LCD display or scratching the display surface.
2.2.1.3 Remove the locking bars from the Recorder enclosure and ensure the gasket is not twisted on the
enclosure.
2.2.1.4 Insert the Recorder rear end first, into the panel opening from the front of the panel as shown in
Figure 2-2. Ensure that the gasket is between the front bezel of the Recorder and the front panel.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Front Panel
Gasket
Figure 2-2 Front Panel Insertion