Chapter 4 Operation

4.1.3 Initialize Database

Three forms of initialization (init) may be performed on the Unit's Profile database: Smart Init, Full Init and Erase
Configuration. The user will not typically see these options unless the unit is being powered up for the first time,
or the unit finds an error in the database. These conditions may be manually invoked using the Hidden menu
option.

4.1.3.1 Smart Init

Smart Initialization verifies that each Profile database block has a correct checksum and (if incorrect) each
parameter in the corrupt block(s) is verified to be within allowable boundaries. Only those parameters in
corrupt blocks found to be outside these boundaries are changed. Bad parameters are always replaced
with pre-defined defaults.

4.1.3.2 Full Init

If the user answers no to "Smart init?", a choice is presented to force the Recorder to do a "Full init?". Full
initialization will completely clear all user-defined parameters, setting them to pre-defined defaults.

4.1.3.3 Erase Config

This option is used on first time power up and will not normally be seen by the user unless the database
has been corrupted or this option is called from the SYSTEM menu. This option will clear all variables to
zero, and should be used to clear memory before programming for the first time, or after upgrading the
firmware. This operation is usually performed at the factory.

4.1.3.4 Init Defaults

The listing below gives the default parameters that are used by both Smart and Full initialization:
Display Block
Power-up Display Mode =
Display Rate =
Time Format =
Language =
Scan Block
Alarm Contact Outputs =
Alarm Checks =
TCBO Test Interval =
Serial Port
Serial Port =
Chart Scales Block
For Each Scale . . .
Scale Type =
Origin =
Decimal Fix =
Scale =
Engineering Units =
Active Scaleset =
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Unit Tag
1 second
American
English
Open on Clear, No Reflash, Failsafe off, Open on ACK
Enabled
300 seconds (5 minutes)
Set to Modbus RTU, 9600, 8bits, parity off, 2 stop bits
Linear
Left
3 places
Low = 0.0 Mid = 50.0 High = 100.0
Set to all spaces (cleared)
Set to scaleset 1
Page 4-2