26. RECHARGE AND SUBSTITUTION OF THE BATTERY
The pressure gauge is powered by one 3.6V 1800mAh RECHARGEABLE Li-Ion battery that guarantees long
battery life and a large number of recharges.
The battery level is signaled by an icon with an indication of the state of charge on 3 levels.
During the charging phase the battery icon will indicate this status with the usual variable indication.
The charging time can last up to about 8 hours. When the charging phase is finished, the battery icon will be
displayed fixed.
BATTERY COMPLETELY DISCHARGE: Measurements made in this state may be altered, therefore an
immediate recharge of the batteries is required.
BATTERY COMPLETELY CHARGED.
Should it be necessary, it is possible to replace the battery with an exact equivalent.
Ask BART for the replacement battery.
Battery replacement is very simple. Simply remove the 2 screws on the rear panel, remove the exhausted
battery and insert the new battery by connecting the dedicated connector.
You can check the voltage in Volts of the battery in the following way (remember that with a fully charged
battery you will have 4.2V while for low battery you mean a value below 3.5V):
+
P0000
3.723
In case of non-use or prolonged storage, we recommend recharging the battery at least once a
month to prevent the battery from discharging completely.
USER GUIDE
Press the keys simultaneously for a few seconds SET and PEAK.
Set the password password 0055 (
The value in Volt of the battery level will be displayed.
Press SET to exit the function and return to the pressure measurement
WARNING
 ) and confirm with SET.
Pag. 18
MO.LABDMM2.560.ENG.R1