Maintenance, Overhaul, and Software Upgrade
While the numbers are flashing, rotate the adjustment knob to change the altitude level as appropriate
17.
(for example, if located at sea level, set altitude to "0 m / 0 ft").
To confirm the selection, press Accept.
18.
Wait 15 seconds to 20 seconds.
19.
Turn the ventilator off.
20.
To access the Software Download/Diagnostics Mode screen, press and hold the Accept button, and then
21.
turn on the ventilator.
Release the Accept button once the ventilator sounds a short beep.
22.
Touch the Start Diagnostics button on the screen.The Diagnostic Data screen will appear, as shown in
23.
Figure 2-27.  
Press the Trig Button on the control panel.
24.
Rotate the adjustment knob to change the value in the display window to "d9."
25.
Press Accept.
26.
Occlude the breathing circuit from the y-piece using a cap (do not use test lung).
27.
To start the exhalation flow sensor calibration (d9), press the Manual Inflation button.
28.
Wait for 30 seconds to 1 minute until the "TSI Air flow:" and "Exhale Flow:" numbers start to increase
29.
gradually and the air flow is delivered from the ventilator. The exhalation flow sensor calibration
requires 6 minutes to 10 minutes to complete (see Figure 2-27.  ).
When the flow delivered from the ventilator stops and "Test Completed" appears on the screen, the
30.
exhalation flow sensor calibration is successfully completed (see Figure 2-27.  ).
Turn the ventilator off.
31.
2-30
Figure 2-27.  Diagnostic Data Screen
Service Manual