Wavelength Accuracy - Jasco V-530 Hardware/Function Manual

Uv/vis spectrophotometer
Hide thumbs Also See for V-530:
Table of Contents

Advertisement

4.4.3 Wavelength Accuracy

Measure the spectra of the D
and read the peak wavelength.
(1) Light only the Deuterium lamp by following items (1) to (3) in Section 4.4.2.
(2) Select [Single Beam] from the [Environment] menu to display the [Single Beam]
screen. (Refer to Section 9-5 of the Operations manual.)
(3) Display the [Parameter] screen and set the following parameters.
<<Parameters for spectrum with an emission line 656.1 nm>>
V-530 measurement parameters
Phtmtrc mode
Response
Start WL
End WL
Data interval
Scan. speed
No. of Cycles
Lower scale
Upper scale
Note:
PMT voltage
<<Parameters for spectrum with an emission line 486.0 nm>>
Start WL
End WL
All other parameter settings are the same as shown above.
(4) Measure the single beam spectrum with the sample chamber empty to obtain the
emission line spectra of the D
and Fig. 4.6.
(5) Wavelength accuracy is normal if the measurement results are within the following
criteria.
<V-530 criteria>
<V-550/560/570 criteria> Within 656.1 0.3 nm
(6) Display the [Others] screen again and return to the original settings (WI & D2).
lamp with emission lines at 656.1 nm and at 486.0 nm
2
V-550/560/570
parameters
Sam
Phtmtrc mode Sam
Fast
Response
658 nm
Band width
654 nm
Start WL
0.1 nm
End WL
40 nm/min
Data interval
1
Scan. speed
0
No. of Cycles
100
Lower scale
Upper scale
HT Voltage
with the instrument. Set accordingly.
varies
488 nm
484 nm
lamp. Measurement examples are shown in Fig 4.5
2
Fast
0.5 nm
658 nm
654 nm
0.025 nm
10 nm/min
1
0
100
300 V
Within 656.1 0.5 nm
Within 486.0 0.5 nm
Within 486.0 0.3 nm
26
measurement