(3) Remounting the sample chamber
a) Locate the sample chamber over the two positioning pins on the main unit and
move it all the way to the back.
b) Ensure that light is incident upon the center of the cell holder. To check it
visually, set the wavelength to 500 nm and place a piece of white paper in front of
the cell holder.
c) Firmly tighten the two fixing screws and close the lid.
4.3

Performance Check: Data Station Type

A performance check should be
instrument performance. If the instrument is not operating properly, conduct checks
and take corrective action after referring to the troubleshooting chart in Section 5. After
taking corrective action, run the performance check again to verify that the instrument
has been restored to normal.
Note:
This section provides a simple description of operating procedures. For
details, refer to the Operations manual.
Baseline flatness
Wavelength repeatability
Wavelength accuracy

4.3.1 Baseline (Abs 0 line) Flatness

This check should be performed more than one hour after turning the instrument on.
(1) Start the [Spectrum Measurement] program from the [Spectra Manager] window.
(2) Click [Measurement]-[Spectrum Parameters...] and set parameters as follows:
V-530 measurement parameters
Photometric Mode Abs
Response
Scanning Speed
Start
End
Data Pitch
Display
No of Cycles
(3) Measure the baseline as follows.
a) Click [Measurement]-[Baseline...]. The [Baseline Correction] dialog box
appears.
b) Click the <Measure...> button to display the [Baseline Measurement] dialog
box.
c) Make sure that the sample chamber is empty and then click the <Start> button.
The baseline is measured and stored in the memory of the spectrophotometer.
run
approx. once a month to ensure the
Medium
200 nm/min
1000 nm
200 nm
1.0 nm
-0.005~+0.005
1
Refer to Section 4.3.1
Refer to Section 4.3.2
Refer to Section 4.3.3
V-550/560/570
parameters
Photometric Mode Abs
Response
Band Width
Scanning Speed
Start
End
Data Pitch
Display
No of Cycles
19
normal
measurement
Medium
2.0 nm
200 nm/min
850 nm
200 nm
1.0 nm
-0.005~+0.005