V-550 Uv/Vis Spectrophotometer Specifications - Jasco V-530 Hardware/Function Manual

Uv/vis spectrophotometer
Hide thumbs Also See for V-530:
Table of Contents

Advertisement

Baseline flatness
Detector
Power requirements
Dimensions and weight
Note:
Photometric repeatability, photometric accuracy and baseline flatness are the
values obtained more than one hour after the "Power" switch is turned ON.
1.2

V-550 UV/VIS Spectrophotometer Specifications

Optical system
Resolution
Light source
Light
source
wavelength
Wavelength range
Wavelength repeatability
Wavelength accuracy
Spectral bandwidth
Wavelength display
Photometric mode
Photometric range
Photometric display
Photometric repeatability
Photometric accuracy
Response
Stray light
±0.001 Abs
(value obtained after baseline correction when
temperature
wavelength: 200 to 1100 nm, response:
Medium,
nm/min.)
Silicon photodiode (S1337)
100, 115, 200, 220, 230, 240 V ±10 %, 130 W
484(W)×435(D)×205(H) mm (excluding
protrusions)
16 kg
Single monochromator
UV/VIS region: 1200 lines/mm plane grating
Czerny-Turner mount
Double beam type
0.1 nm
Deuterium lamp: 190 to 350 nm
Halogen lamp: 330 to 900 nm
changeover
Any wavelength between 330 and 350 nm can
be selected.
190 to 900 nm
±0.1 nm (at a spectral bandwidth of 0.5 nm)
±0.3 nm (at a spectral bandwidth of 0.5 nm)
0.1, 0.2, 0.5, 1, 2, 5, 10 nm
L2, L5, L10 nm (low stray-light mode)
In increments of 0.1 nm
Abs, %T. %R, Sample, Reference
-2 to 3 Abs
0.0 to 200.0 %T
-2 to 5 Abs
Four digits and (-) symbol
Displays up to 000.0 %T and 0.000 Abs
±0.001 Abs (0 to 0.5 Abs)
±0.002 Abs (0.5 to 1 Abs)
±0.002 Abs (0 to 0.5 Abs)
±0.004 Abs (0.5 to 1 Abs)
±0.3 %T
Note:
Quick, Fast, Medium, Slow
0.015 % (220 nm: NaI 10g/L aqueous solution;
340 nm: NaNO
variation
and
wavelength
Tested with NIST SRM 930D
50g/L aqueous solution)
2
2
is
within
5°C,
scanning:
100

Advertisement

Table of Contents
loading

This manual is also suitable for:

V-550V-560V-570

Table of Contents