HP 53150A Series Programming Manual page 35

Microwave frequency counter
Table of Contents

Advertisement

Chapter 2 Command Summary
HP 53150A/151A/152A Command Summary
Table 2-2. HP 53150A/151A/152A SCPI Command Summary (Continued)
Keyword/Syntax
Parameter Form
:MEMory
:CLEAr[:NAME]
<name>
:DATA
<name>, <data>
:DATA?
<name>
:NSTates?
:READ
[:SENSe]
:AVERage
[:STATe]
<Boolean>
:COUNt
<numeric_value>
:CORRection
:CSET
:SELect
<character_data>
CORR1 | CORR2 . . . CORR9
:STATe
<Boolean>
:DATA?
<data_handle>
"[XNONe]FREQuency [1 | 2]"
"[XNONe]POWer [2]"
:FILTer
:FM
:AUTO
<Boolean>
Programming Guide
Std/New Comments
Std
Subsystem. Manages instrument memory.
Std
Event; no query. Restores the frequency
values in the named correction profile to the
default values and sets all loss values to
zero.
Std
Stores data in the named correction profile.
Std
Queries the data in the named correction
profile.
Std
Query only. Returns the number of available
*SAV/*RCL states in the instrument.
Std
See Measurement Instructions in this table.
Std
Subsystem setup commands.
New
Subtree. Configures the averaging function.
New
Turns averaging ON and OFF.
New
Specifies the number of measurements to
combine when AVERage:STATe is ON.
Std
Subtree. Configures the power-correction
function.
Std
Selects a power-correction profile.
Std
When STATe is ON, power measurements
are modified according to the data in the
correction profile selected with :SELect.
Std
Query only. Returns the current
measurement result data of the SENSe
subsystem.
Std
Frequency on channel 1 or 2.
Std
Power on channel 2.
Std
Subtree. Controls the use of filtering
routines in the instrument.
New
Turns automatic FM compensation ON or
OFF.
2-15

Advertisement

Table of Contents
loading

This manual is also suitable for:

53152a series53151a series

Table of Contents