HP 53150A Series Programming Manual page 125

Microwave frequency counter
Table of Contents

Advertisement

Chapter 4 Command Reference
[:SENSe] Subsystem
[:SENSe]:FREQuency:OFFSet . . . <numeric_value>[<frequency unit>]
COMMAND
Sets a reference frequency for all other absolute frequency settings in the
instrument.
The acceptable range for the <numeric_value> parameter is 0 to 50 GHz.
RANGE
The offset frequency can be specified in Hz, KHz, or MHz only.
UNITS
This command does not affect the hardware settings of the
COMMENTS
instrument. It affects only the entered and displayed frequencies.
The Counter accepts only six digits of resolution for an offset entry
through this command. For example, the command
FREQ:OFFSet 12345.678912MHz results in an offset of
approximately 12.3456 GHz.
The coupling equation for this command is:
Entered | Displayed frequency = (Hardware frequency ) + offset.
Freq Offset
RELATED
FRONT-PANEL
KEYS
[:SENSe]:FREQuency:OFFSet:STATe . . . <Boolean>
COMMAND
When FREQuency:OFFSet is ON, the frequency measurement results are
modified by the setting of FREQ:OFFset before being displayed or
reported (in response to a query).
Freq Offset
RELATED
FRONT-PANEL
KEYS
Programming Guide
4-25