HP 53150A Series Programming Manual page 159

Microwave frequency counter
Table of Contents

Advertisement

Chapter 4 Command Reference
Common Commands
*TST?
(Self-Test Query)
*TST?
COMMAND
This query causes an internal self-test; the response indicates whether
any errors were detected.
When the self-test fails, error -330 is generated, and other messages
indicating specific failures are also placed in the error/event queue.
QUERY RESPONSE
Table 4-4. Self-Test Error Values
Bit
Test Failed
0 Band 1 Signal Path
1 Band 1 Threshold
2 Band 2 RF Threshold
3 Band 2 IF Through Threshold
4 Band 2 IF Heterodyne Threshold
5 Heterodyne Path
6 Through Path
7 VCO
8 Counter Control FPGA
9 Power Measurement Hardware
10-15 Unused
16 –12 V
17 –5 V
18 +12 V
Programming Guide
Numeric data transferred as ACSII bytes in <NR1> format.
A response value of zero indicates the self-test has completed with no
errors detected, while a non-zero value indicates the self-test was not
completed or was completed with errors detected. The test failures
that can be detected and their corresponding bit numbers are shown
below:
The decimal weight of each bit is 2
Bit
Test Failed
19 ROM
20 Unused
21 Front Panel Hardware
22 HP-IB
23 EEPROM Instrument Configuration Data
24 EEPROM Service Data
25 EEPROM Saved User Settings Data
26 EEPROM Power Calibration Data
27 EEPROM Write
28 EEPROM Power Correction Data
29 ADC
30 Over Temperature
31 Unused
n
, where n is the bit number.
4-59

Advertisement

Table of Contents
loading

This manual is also suitable for:

53152a series53151a series

Table of Contents